ASA Glass/ Metal Tube Flowmeter
Electromagnetic Flowmeter
Electromagnetic Flowmeter

Electromagnetic Flowmeter

產品特色

This new model, at high performance solutions, both as hardware and software, can be defined as an HI-TEC line, addressed to very demanding customers from the point of view of performance and precision.

Output signal is both analogical and digital; and qualitatively the instrument is very performing, having accuracy < +-0,5% of reading and repeatability of 0,1% r.v.

ASAMAG is also good to be used with any liquids with conductivity better than 5 μS/cm; they are insensitive to temperature and pressure variations, density or viscosity of process fluid.

ASAMAG advantages are:




  • maximum control (also in remote). ASA can provide also a serial interface software usable by PC and the port RS232 (or RS485) of the converter, making possible the visualization and the modification of process parameters.

  • all ASAMAG are equipped with Sensoreprom unit that contains all calibration data of sensor tube and all the settings of converter for the lifetime of the product. Fabrication parameters and instrument size are recordered inside the Sensoreprom together with the working parameters set by customer. In the event of tube substitution, Sensoreprom make it very easy and rapid.

  • great flexibility is intended as the chance to have compact or remote version with the same measuring tube and converter.



For a complete electromagnetic flowmeter you have to combine ASAMAG converter with sensor tube of  Flanged, Wafer or DIN-Clamp-Gas model.


Quantity

Product specifications

Measuring Range:

1:10(depends on process condition)

Accuracy:

± 2% f.s.v. standard; better accuracy optional

Repeatability:

< ± 0.25% of actual read value

Max pressure:

40 bar (higher on request)

Advised Fluid Temperature:

0-90°C with Buna O-Rings

0-120°C with Viton O-Rings

Process connection:

female gas screwed UNI 338 and flanged UNI EN-1092-1 standard.

Other types of connections screwed or flanged (ANSI, DIN, etc.) optional.

Wetted Parts material:

Carbon streel standard.

AISI 316L stainless steel, Hastelloy B and C, Monel, Titanum, PVDF optional

Meter Tube material:

borosilicate glass

Float material:

Spherical float: glass, AISI 316 stainless steel, tungsten carbide, standard

Cone float: AISI 316L stainless steel, aluminium, standard

Float Stops: AISI 316L stainless steel standard

Housing Material:

Carbon steel painted with epoxy resin standard.

AISI 304 stainless steel

Optional

Flange Material:

carbon steel (painted) standard.

AISI 316L stainless steel, optional.

 

Measuring Range:

0.3m/s~12m/s

Accuracy:

± 0.5% r.v..

Repeatability:

< ± 0.1% r.v.

Max pressure:

10 bar

Internal lining:

PTFE, Hard Rubber and DIFLEX

Electrodes:

AISI 316L(standard)

Titanium, Hastelloy “C” , Tantalum(optional)

Measuring Tube material:

AISI 304 coated

Board cover:

die casted aluminium epoxy painted RAL 7001 (AISI 304 as option)

Tube mechanical protection:

IP65(standard)

IP67/68 on request:

Temperature:

(Electronic board)

PTFE -10° + 75°C (176°F)

EBANITE -10° + 75°C (176°F)

DIFLEX -10° + 75°C (176°F)

(Measuring tube)

PTFE -30° + 130°C (266°F)

EBANITE -10° + 80°C (176°F)

DIFLEX -10° + 160°C (320°F)

Connections:

WAFER

FLANGE

TRICLAMP

Transmitter:

(Output)

0/4-20mA(max load 1.400 Ω)

1x programmable pulse output (max 10 KHz), auto pulse length

1x output open collector to be used as flow or flow-direction alarm, totalization alarm or error alert

2x relays for system errors monitoring, dosing, flow/flow-direction / totalization alarm

(Communication protocol)

Hart, RS485, RS232 on request

(Power supply)

From 20 to 55 Vdc and from 17 to 45 Vac (24V version)

From 90 to 250 Vac (220V version), 50/60Hz

(Transmitter Housing)

Die cast aluminium IP 67; IP 68 on request

(Environmental temperature)

-10° + 75° C